Skip to content
Open Data Finder
data.gov.au

RV Investigator Ultrafine Aerosol Volatility and Hygroscopicity (VH-TDMA) - Selected Voyages (2016 onwards)

Published by CSIRO Marlin Data Catalogue

About this dataset

The VHTDMA was constructed at QUT for measurement of ultrafine aerosol volatility and hygroscopicity. It uses electrostatic classifiers to select particles of a particular electrical mobility diameter, conditions them (using a thermodenuder and humidifier) and re-measures the particle diameter. The VH-TDMA is made up of three purpose built electrostatic classifiers and two condensation particle counters (CPCs). One TSI 3010 CPC and one TSI 3787 CPC. The pre-classifier (used to select particles of a particular dry diameter) most commonly cycled between measuring 40, 100 and 150 nm diameter particles during IN2016_v02. For short periods other particle diameters were selected (20, 30, 60,80 nm), to examine features in the ambient size distribution. The thermodenuder was set to 250 °C for the majority of IN2016_v02, with short periods at other temperatures (150°C, 200°C). The relative humidity of the H-SMPS was set to 90% RH for most of the voyage, however it was periodically lowered for size calibration scans and to measure the deliquescence RH for ambient particles. The sample was passed through a nafion drier and inlet RH logged to ensure that dry particle diameter was selected i.e. RH < 30%, the efflouresence RH for common marine salts.

RV Investigator Ultrafine Aerosol Volatility and Hygroscopicity (VH-TDMA) - Selected Vo… · Open Data Finder